Abbott initiates voluntary recall for certain FreeStyle Libre 2 Sensors in UK

Abbott is initiating a voluntary recall of certain FreeStyle Libre 2 sensors from lot KTP005061. No other Abbott diabetes products are affected. ​

Problem / Issue

Abbott has recently identified that certain FreeStyle Libre 2 sensors from lot KTP005061 may provide erroneously high glucose readings. For example, your sensor glucose reading may be high and out of your target range while your actual glucose levels may be below or within target range.​

Potential Harms ​

If undetected, erroneously high glucose readings can pose a potential health risk for people living with diabetes. ​

Erroneously high glucose readings can lead to incorrect treatment decisions, such as taking insulin when not required ​

Actions for Users ​

To determine if your current sensor or any unused sensors you have are affected, please visit www.FreeStyleConfirm.com and enter your sensor serial number. ​

If you are currently using the FreeStyle Libre 2 system and are wearing an affected sensor, please immediately discontinue use. If you have symptoms that do not match the Sensor glucose reading, or suspect your reading may be inaccurate, check the reading by conducting a fingerstick test using a blood glucose meter. ​

You will be sent a replacement for any affected sensor(s). Abbott will send you a mailing kit to return any affected sensor(s) to them. ​

While waiting for your replacement Sensor, you can continue to use sensors that are not affected and can use the built-in blood glucose meter in your FreeStyle Libre 2 reader to check your glucose at any time. ​

Abbott sincerely apologizes for any inconvenience this has caused. ​

If you have questions, please contact Abbott Customer Service at 0800 170 1177. You can call us Monday to Friday between 8:00 AM and 8:00 PM, Saturday 9:00 AM to 5:00 PM (excluding Bank Holidays).

See here for details on the Abbott website.